Level measurement

113 Selection according to application

Slim, narrow silos, vessels

• Filling via mechanical or pneumatic conveyance •Fluidization possible •

H

D

Contact

B

Our proposal Guided radar Levelflex

Electromechanical level system Silopilot

FMP56

FMP57

FMM50

FMM20

• Unaffected by silo geometries and the shape of the angled surfaces • Unaffected by the density of bulk solids, temperature, humidity and filling noise • • Heartbeat Technology

• Unaffected by low density of bulk solids and DC value • Easy installation

2-wire (HART®, PA, FF), 4-wire HART®

4-wire, 4-20mA, relay ±2.5cm/±1" (FMM20), ±5cm/±2" (FMM50)

–40 to +150°C/–40 to +302°F –1 to +16bar/–14.5 to +232psi 1.4

–20 to +230°C/–4 to +446°F +0.8 to +3bar/+11.6 to +44psi — DN 100 PN 16 (hole size) 70m/230ft (special design up to 90m/295ft)

45m/148ft

• Abrasive, grained, lumpy products

radar, ultrasonics

•Risk of

radar, ultrasonics

weight being buried

• Max. tensile forces on the rope = 35kN (observe ceiling load) • Extreme build-up formation on the probe •

radar, ultrasonics, electrom. level system radar with purge air, ultrasonics radar, electrom. level system ultrasonics, electrom. level system radar, electrom. level system electrom. level system

•Strong

radar, ultrasonics

mechanical wear to be expected • Measurement during filling

guided radar, radar, ultrasonics

• Measuring range > 45m/ 148ft powdery products •

Please note: Guided radar continued on Page 122

Please note: Electrom. level system continued on Page 132

Made with FlippingBook Digital Proposal Maker