127 Instrument selection within the measuring principle
Sensor alignment • Angled surfaces are formed in silos for bulk solids. These cause the ultrasonic signal to be laterally reflected which can lead to a reduced signal intensity Remedial measures: The sensors should be aligned as vertically as possible in relation to the product surface This is facilitated by the FAU40 alignment device or the assembly bracket
Advantages • Non-contact, maintenance-free measurement • density, etc. • Calibration without filling or discharging • Self-cleaning effect of sensors due to moved sensor diaphragm • Separate instrumentation options in rough ambient conditions • Cost-effective instrumentation for silo farms with FMU95 multichannel system
C
FAU40 for Prosonic
Installation with assembly bracket for Prosonic
Range calculation and sensor selection Prosonic M FMU4x
R [m/ft]
20/65
FMU44
10/32
Example (for FMU43): • Product surface hard, rough: • Low dust generation: • Small quantities of filling curtain in the detection range: Total: range approx. 7m/23ft from diagram
FMU42
7/23 8/26
+40dB +5dB +5dB +50dB
FMU41
5/16
FMU40
1/3.2
10 20 30 40 50 60 70 80
A [dB]
Made with FlippingBook Digital Proposal Maker