135 Instrument selection within the measuring principle

Installation instructions – electromechanical level system

Installation • Not in the filling curtain or in the area of collapsing product accumulation on walls • Measuring point as close to the center of the slope as possible • The sensing weight may neither sink into the product nor slide off the angled surface during the measuring operation • Max. angle of inclination 2° Weather protection cover • Always recommended for installation outside (solar radiation and rain) Compressed air connection • Already integrated and the penetration of dust can be avoided in case of strong dust generation Tank baffles • The measurement section should not pass baffles and struts at too close a distance. The measuring tape must not touch any baffles and struts

C

*1 Accumulation (product build-up on the wall of the vessel) *2 Choose a measuring point located approximately in the middle of the slope