46 Continuous level measurement in liquids

4. Instrument selection within the measuring principle

Measuring range in dependence on the type of tank

Process conditions and medium for Micropilot FMR51/FMR52

Tank with single-stage propeller agitator Turbulent surface, single-stage agitator

C

Horn/antenna diameter

FMR51 FMR52

40mm/1.5"

50mm/2" 50mm/2"

80mm/3" 80mm/3"

100mm/4"

Measuring range in m/ft

Media group A: B: C: D:

B C D

C D

B C D

B C D

2/ 6.6

2/ 6.6

2.5/ 8.2

Standard: Max. measuring range = 40m/131ft With application package “Advanced dynamics“: Max. measuring range = 70m/229ft Min. measuring range = 5m/16ft

3/ 9.8

3/ 9.8

4/ 13 5/ 16

5/ 16

5/ 16

5/ 16

8/ 26

7.5/ 25

8/ 26

10/ 32

10/ 32

12/ 39

15/ 49

15/ 49

20/ 65