Level measurement

22 Continuous level measurement in liquids

3. Selection of the measuring principle according to the application

Non-contact

B

Our proposal Radar Micropilot

Ultrasonics Prosonic

(separated)

(compact)

FMU90

FDU9x

FMR5x

FMR6xB

FMU4x

Advantages

• Non-contact and unaffected by head pressures

• High resistance • Self-cleaning effect of sensors • Integrated alarm/point level relay • Fast measuring frequency (4-wire)

• Universally useable due to •Flexible measuring range • Changing, highly viscous or •Remote access via Bluetooth® • Heartbeat Technology

aggressive media (100% PTFE)

Technical data • Connection •Accuracy •Process temperature •Process pressure •Process connection

2-wire (HART®, PA, FF, Ethernet-APL) ±1mm/±0.04" –196 to +450°C/–321 to +842°F –1 to +160bar/–14.5 to +2,320psi Threads, flanges (DIN, ANSI, JIS), hygienic connections 80m/262ft

2-/4-wire (HART®, DP) ±2mm/±0.08", +0.2% of the distance –40 to +105°C/–40 to +221°F +0.7 to +4bar/+10 to +58psi Threads, Tri-Clamp, flanges (DIN, ANSI, JIS)

•Maximum

25m/82ft

measuring range

Application limits

•Strong formation of foam •Many obstacles • Low DC value

• Strong formation of foam • Vapor pressure • Many obstacles • Fast temperature changes •Strong turbulences

hydrostatics

radar hydrostatics radar

hydrostatics

•Extreme

turbulences

hydrostatics

Please note: Radar continued on Page 34

Please note: Ultrasonics continued on Page 74

Made with FlippingBook Digital Proposal Maker