Instrument selection within the measuring principle
75
Vapor pressure of the medium (20°C/68°F) The vapor pressure of the medium at 20°C/68°F is an indication for the accuracy of ultrasonic level measurement. If the vapor pressure at 20°C/68°F is lower than 50mbar/0.73psi, ultrasonic measurement is recommended. If the vapor pressure at 20°C/68°F is above 50mbar/0.73psi, the accuracy of the measurement will be affected. To achieve the highest accuracy results, radar level measurement is recommended.
Advantages • Non-contact, maintenance-free measurement • etc. • Calibration without filling or discharging • Self-cleaning effect due to vibrating sensor diaphragm
Vapor pressure
Examples
Water, water solutions, water-solids solutions, dilute acids (hydrochloric
C
(20°C/68°F)
(20°C/68°F)
Range calculation and selection of sensor for Prosonic M FMU4x and FMU30
Example (for FMU41): • Very turbulent surface: 20dB • medium surface approx. 60°C/140°F: 15dB • Total: 35dB range approx. 6m/20ft
R[m/ft]
20/65
FMU44
10/32
FMU42
8/26
FMU41
Example (for FMU30 2" sensor): • Strong turbulence surface:
5/16 6/20
approx. 20dB
FMU40
• No dust formation: • Filling curtain in detection range:
0dB
1/3.2
10dB
10 20 30 40 50 60 70 80 A [dB]
Total: approx. 30dB range approx. 7.8m/26ft from diagram